simple-spectro

Very cheap, autonomous and precise spectrophotometer for DIY biology and chemistry.

This project is maintained by Hackuarium

Hacking of the case values

Problems were encountered because too much light was going through the sample. An additional part is tested, which reduces the slit through which the light goes.

Empty tank without additional part

measure # R G B
1 353348 317967 59192
2 353213 317995 59189
3 353089 317944 59185
4 353028 317978 59181
5 353032 317975 59194
6 352949 317994 59184
7 352967 317949 59191
8 352905 317962 59184
9 352938 317999 59179
10 352865 318019 59180

#

Tank H2O without additional part

measure # R G B
1 472480 421630 59171
2 472415 421713 59166
3 472383 421649 59171
4 472367 421823 59173
5 472356 421606 59177
6 472217 421674 59171
7 472232 421680 59171
8 472146 421702 59167
9 472221 421524 59170
10 472161 421437 59170

#

Empty tank with additional part

measure # R G B
1 327342 322453 397693
2 327361 322496 397858
3 327468 322470 397821
4 327434 322432 397863
5 327419 322515 398088
6 327444 322432 397994
7 327421 322459 398066
8 327478 322446 398031
9 327403 322425 398044
10 327452 322447 398030

#

Tank H2O with additional part

measure # R G B
1 414183 422649 50470
2 414031 422679 50478
3 414144 422622 50485
4 414044 422657 50464
5 414163 422679 50532
6 414049 422611 50522
7 414145 422619 50530
8 414082 422573 50538
9 414129 422544 50544
10 414115 422604 50553

#

Orientation with additional part

measure # R G B
1 61773 141232 177960
2 60593 142047 175869
3 61773 141232 177960
4 60593 142047 175869
5 60549 138233 179474
6 61081 141630 177519
7 61070 141654 177439
8 61194 140962 177646
9 61172 140918 177628
10 61732 141052 178012

#

Orientation curve without additional part

measure # R G B
1 88000 258507 182288
2 93625 256857 180842
3 77947 212538 180771
4 95924 281735 182485
5 95897 281571 182459
6 99752 269188 182928
7 98857 272064 182035
8 83031 235556 181643
9 83021 235429 181570
10 96018 262922 181537

#